Vacuum Products

KVC as a vacuum measuring instrument specialized company aiming to become a leading company in 21stcentury control and measurement industry

KM700N Digital Manometer Controller

KM700N Digital Manometer Controller

Manometer Gauge Controller

Measurement Range
1 to 1500Torr

Display unit
Torr, KPa

Application Sensor
KM70S : Manometer Sensor

Optional : Interface
RS485 ASCⅡ, RS485 ModBus RTU
RS232 ASCⅡ, RS232 ModBus RTU
Specitification downloadDownload
Dimension downloadDownload
Manual downloadDownload

​KM700N Digital Manometer Controller measures vacuum according to the user's purpose and utilizes the output to control various vacuum systems and pressure. 

 

 

Key Feature 

 

-KM700N vacuum gauge controller is a pannel mounting type in DIN specification (96mm x 96mm) size.

 

-Key operation is easy, so it is convenient to set it up.

 

-High luminance chip LED module used for easy management as the display is large and clear.

 

-KM7000N is compact design and accuracy, stability, reliability, reproducibility,

 

 durability will help with company quality and productivity.

 

-Made of all-metal, there is no problem with RF interference.

 

-CE-certified product.

 

 

 

 

Need help?

Contact an Application Specialist by sending an email to KVC or Call 032-234-3030

 

※ ​How to choose the Best Product


 

Product Specification 

- Controller Specification

 

 Measuring Range for Air and N2

 1 ~ 1500 Torr

 Operating Temperature

 0 ~ 60 ℃

 Storage Temperature

 - 40 ~ 70 ℃

 Case Material

 Stainless steel

 Compliance  

 CE ( EMC ) 

 Display 

 4 Digit LED , Status LED

 Set Point Relays

 2 Point SPST ( 5A / 250VAC, 5A / 30VDC )

 Analog Output

 Log / Lin scale ( 0 ~ 10 Vdc ) Selectable

 Digital Interface

 RS232 / 485 ASCII , RS232 / 485 MODBUS

 Power Required

 AC 85~260V 50/60Hz 20VA

 Weight

 630 g

 Mounting 

 Pannel mounting ( DIN 96 * 96 mm )

 


- Sensor Specification

 

 Sensor Type

 Manometer KVC70S

 Sensor Materials

 Piezo Resistive silicon Sensor with sus housing 

 Internal Volume

 5 cm

 Gauge Bakeout Temperature

 150 ℃ Maximum, NON - operating

 Case Materials

 304 Stainless steel, borosilicate glass, Kovar

 ​


※ Do not use in explosive atmospheres or flammable gases or vapors.

 Do not use it to measure the pressure of explosive gas, combustible gas, or mixed gas.