Vacuum Products

KVC as a vacuum measuring instrument specialized company aiming to become a leading company in 21stcentury control and measurement industry

KVC450 Compact SHIM Module Vacuum Gauge

KVC450 Compact SHIM Module Vacuum Gauge

Convection Vacuum Gauge

Measurement Range
1.0E~3Torr to 1000Torr

Display unit
Torr, Pa

Application Sensor
KVC450 SHIM Vacuum Gauge

Optional : Interface
RS485 ASCⅡ, RS485 ModBus RTU

Specitification downloadDownload
Dimension downloadDownload
Manual downloadDownload

(Also Available as Vacuum Switch)

KVC450 Compact SHIM Module Vacuum Gauge features a compact design, which reduces installation costs, makes installation convenient, and is an ideal product for local control.

Various vacuum systems and pressure control are possible by utilizing the output.



Key Feature 

 

- KVC450 has 2 setpoints and can be used as a vacuum switch.

 

-Compact design and accuracy. stability. reliability. reproducibility. durability will help with company quality and productivity.

 

-Made of all-metal, there is no problem with RF interference.

 

-CE-certified product.

 

 

 

Need help?

Contact an Applications specialist by sending an email to KVC or Call 032-234-3030


※ ​How to choose the Best Product

 

 

Product Specification 

- Controller Specification

 

 Measuring Range for Air and N2

 1.0E-3 ~ 1000 Torr

 Operating Temperature

 0 ~ 60 ℃

 Storage Temperature

 - 40 ~ 70 ℃

 Case Material

 Aluminum

 Compliance 

 CE ( EMC )

 Display 

 4 Digit LED, Status LED

 Set Point Relays

 2 Point SPST ( 5A / 30VDC, 5A / 250VAC )

 Analog Output

 Log scale ( -3 ~ +3 Vdc ) Adjustable Bias

 Lin Scale ( 0 ~ 10 Vdc ) Fixed

 Digital Interface

 485 ASCII / 485 MODBUS

 Power Required

 DC 24V / 200 mA 

 Weight

 450g ( With sensor ) 

 Mounting 

 Field Flange Mounting

 


- Sensor Specification

 

 Sensor Type

 KVC450 Sensor

 Sensor Materials

 Gold Plated tungsten or Platinum  

 Internal Volume

 45 cm

 Gauge Bakeout Temperature

 150 ℃ Maximum, NON - operating

 Case Materials

 304 Stainless steel, borosilicate glass, Kovar

 ​

 


※ Do not use in explosive atmospheres or flammable gases or vapors.

 Do not use it to measure the pressure of explosive gas, combustible gas, or mixed gas.